Epson C82302/04 (Serial I/F) Manuel d'utilisateur Page 60

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 294
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 59
2-4
Power Requirements
CAUTION
Before the power is turned on for the first time at
installation, be sure that the unit has been upright for at least four
hours at normal indoor temperatures (15 degrees C to 30
degrees C).
The GeneAmp PCR System 9600 operates on one of the
following input voltages:
* For 220-volt operation with instrument Part No. N801-0002, a
conversion kit is required and included
(Applied Biosystems Part No. N801-0110).
** For 60-Hertz operation with instrument Part No. N801-0003, a
conversion kit is required and included
(Applied Biosystems Part No. N801-0111).
120 VAC and 100 VAC units are supplied with a three-terminal
plug. One of the terminals is an earth ground. To protect yourself
against shock hazards, the electrical outlet
must
be a properly
wired three-terminal outlet conforming to National Electric Code.
Do not use an adapter to a two-terminal outlet.
Part No.
Voltage
(volts)
Freq.
(Hz)
Maximum
Operating
Current
(amps) Power (watts)
N801-0001 120 60 10 850
N801-0002 230/240 50 5 850
220* 50 5 850
N801-0003 100 50 12 850
100 60** 12 850
Vue de la page 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 293 294

Commentaires sur ces manuels

Pas de commentaire