7-52
Running the Chiller Test
Diagnostic Test #3 is the Chiller Test, which verifies the
performance of the refrigeration system by calculating the cooling
rate of the sample block.
CAUTION
Before you begin this test, make sure you place an
empty MicroAmp Tray (MicroAmp Tray/Retainer Set, Part No.
N801-0530) on the sample block, then slide the heated cover
forward and turn the cover knob clockwise until the white mark on
the knob lines up with the white mark on the cover.
To run the Chiller Test:
1
Select DIAG on the Utilities menu as described on page 7-3.
The following display appears:
2
Select Diagnostic Test #3 by pressing
3
, then pressing
ENTER
, or
by pressing the
STEP
key twice.
The following display appears:
3
Press
ENTER
to start the test.
Enter Diag Test #1
REVIEW HISTORY FILE
Enter Diag Test #3
CHILLER TEST