8-16
10
After you measure the second temperature of well A1, turn the
cover knob completely counterclockwise, then slide the heated
cover back.
11
Move the probe assembly to wells A4 and A5 with the dummy probe
in A5.
12
Slide the heated cover forward and turn the cover knob clockwise
until the white marks are aligned.
13
Repeat the measurements on wells A4, A8, A12, C1, C4, C8, C12,
E1, E4, E8, and E12, H1, H4, H8, and H12. Make sure you place
the measuring cone of the probe assembly into these wells and the
dummy probe into adjacent wells.
Figure 8-7
Location of wells used in uniformity test.